SPECIFICATIONS
Wavelength Range |
190.0-1100.0nm |
Display Wavelength |
0.1nm step |
Selectable Wavelength |
0.1nm step (1 nm step in spectrum mode) |
Scan Speed |
Wavelength change: approximately 3800nm/min. Scan: approximately 24-1400nm/min |
Light source change |
Selectable from the following 3 types: Auto change with wavelength Selectable wavelength: 295nm-364nm: 1nm interval Recommended wavelength : 340nm Halogen lamp only Deuterium (D2) lamp only |
Measurement method |
Single beam measurement |
Light Source |
Auto correction with the computer memory 20W Halogen lamp (long-life 2000 hours) Deuterium lamp ( socket type) Auto adjustment for maximum sensitivity |
Monochromator |
Incorporates aberration-correcting concave blazed holographic grating |
Detector |
Silicon photodiode |
Spectrum Bandwith |
5nm |
Wavelength Accuracy |
+/- 1.0nm |
Wavelength repeatability |
+/- 0.3nm |
Stray Light |
Less than 0.05% (220.0nm Nal, 340.0nm NaNO2 & UV39) |
Photometric Range |
Absorbance: -0.3-3.0 Abs Transmittance: 0.0-200% |
Recording Range |
Absorbance: -3.99-3.99 Abs Transmittance: -399-399% |
Photometric accuracy |
+/- 0.005 Abs (at 1.0 Abs) +/- 0.003 Abs (at 0.5 Abs) |
Photometric repeatability |
+/- 0.002 Abs (at 1.0 Abs) |
Drift |
less than +/- 0.001Abs/h (after 2hr warm up) |
Baseline Flatness |
+/- 0.010 Abs (after 1 hr warm-up, at 1100200nm |
Noise |
less than 0.002 Abs, Peak to Peak less than 0.0005Abs. RMS |
Sample Compartment |
Interior dimensions W110.0xD230.0xH105.0nm (Partial Depth: 155.0nm) 2-screw port for optional accessory installation |
Display |
6-inch LCD (320 x 240 dot) with CFD lighting with contrast adjustment |
Power Supply |
100-120V 50/60Hz 160VA 220-240V 50/60Hz 160VA |
Dimensions |
W416 x D379 x H274mm |
Weight |
11kg |
Ambient Temperature |
15-35 Degree C |
Ambient Humidity |
45-80%, less than 70% if over 30 Degree C |